Double-Sided Laser Heating System at Hpcat for in Situ X-Ray Diffraction at High Pressures and High Temperatures

Y. Meng,G. Shen,H. K. Mao
DOI: https://doi.org/10.1088/0953-8984/18/25/s17
2006-01-01
Journal of Physics Condensed Matter
Abstract:An overview of a YLF:Nd laser heating system at the undulator x-ray diffraction station ( 16ID-B) of the high-pressure collaborative access team ( HPCAT) of the Advanced Photon Source is presented. Based on the double-sided laser heating technique, the system is designed with considerable effort on the mechanical and optical stabilities, features for user-friendly operation, and the capability of accommodating diamond anvil cells of various heights up to 68 mm. This system has been used for x-ray diffraction studies of a wide range of materials to over 150 GPa and above 3000 K. Applying the laser heating technique to radial x-ray diffraction studies at simultaneous high-pressure and high-temperature ( PT) conditions requires heating to be conducted at variable angles relative to the x-ray direction. A rotation laser heating design is discussed.
What problem does this paper attempt to address?