Rapid, Sensitive, and Multiplexed On-Chip Optical Sensors for Micro-Gas Chromatography.

Karthik Reddy,Yunbo Guo,Jing Liu,Wonsuk Lee,Maung Kyaw Khaing Oo,Xudong Fan
DOI: https://doi.org/10.1039/c2lc20922e
IF: 7.517
2012-01-01
Lab on a Chip
Abstract:We developed and characterized a rapid, sensitive and integrated optical vapor sensor array for micro-gas chromatography (mGC) applications. The sensor is based on the Fabry-Perot (FP) interferometer formed by a micrometre-thin vapor-sensitive polymer layer coated on a silicon wafer. The thickness and the refractive index of the polymer vary in response to the vapor analyte, resulting in a change in the reflected intensity of the laser impinged on the sensor. In our study, four different polymers were coated on four wells pre-etched on a silicon wafer to form a spatially separated sensor array. A CMOS imager was employed to simultaneously monitor the polymers' response, thus enabling multiplexed detection of a vapor analyte passing through the GC column. A sub-second detection time was demonstrated. In addition, a sub-picogram detection limit was achieved, representing orders of magnitude improvement over the on-chip vapor sensors previously reported.
What problem does this paper attempt to address?