Integrated interferometers as a new platform for low cost gas chromatography detection

Pomme Hirschauer,Benoît Paris,Sonia Messaoudene,Maryse Fournier,Bertrand Bourlon,Yanxia Hou,Florence Ricoul,Loïc Laplatine
DOI: https://doi.org/10.1016/j.talanta.2024.126659
IF: 6.1
2025-01-01
Talanta
Abstract:Gas chromatography is a reference method for gas analysis. As part of efforts to miniaturize gas chromatography systems, the miniaturization of detectors is essential. In this work, we report a new integrated photonic platform for gas chromatography analyte detection. The fabricated silicon die integrates Mach-Zehnder interferometers into low dead volume microfluidic channels, with coherent cost-effective detection scheme with a fixed 850 nm wavelength laser. A proof of concept is demonstrated with the separation and detection of three volatile organic compounds: heptane, octane, and toluene. Peaks' widths at half height range from 1 to 5 s. Peaks are very well resolved by our system, which acquires more than 100 points per second. From a heptane dilution range, we evaluate the limit of detection of our system to be the headspace of a 0.26 % heptane concentration solution. To our knowledge, these are the first integrated Mach-Zehnder interferometers reported for gas chromatography detection. This work could open new strategies for fast low cost and low limit of detection specific gas chromatography silicon micro-detectors.
What problem does this paper attempt to address?