[Strength of Bi-Layer Dental Ceramic Evaluated by MSP Testing].

Yi-Hong Liu,Yan-Hong Zhu,Qi-Huang Deng
DOI: https://doi.org/10.3969/j.issn.1671-167x.2011.06.021
2011-01-01
Abstract:OBJECTIVE:To evaluate strength of bi-layer alumina/Y-TZP specimens, compared with 3-point bending test and to estimate the reliability of MSP testing used in dental ceramics.METHODS:The single-layer/bi-layer alumina/Y-TZP specimens were produced by the clinical methods. Strength of these specimens was determined by MSP testing and 3-point bending test respectively. Fractured pieces of each bi-layer specimen were collected and studied with optical microscope and SEM.RESULTS:The MSP fracture strength was lower than bending strength in all the specimens. The strength of single-layer Y-TZP was the highest and the strength of bi-layer samples was lower than that of the single-layer samples respectively. Delamination and interface destruction were found in bi-layer alumina specimens, and not in bi-layer Y-TZP specimens.CONCLUSION:MSP testing is a convenient and feasible method to evaluate mechanical properties of dental ceramics.
What problem does this paper attempt to address?