Electron streaming dose measurements and calculations on a 1.5 T MR‐Linac
Elizabeth Patterson,Marcus Powers,Peter E. Metcalfe,Dean Cutajar,Bradley M. Oborn,John A. Baines
DOI: https://doi.org/10.1002/acm2.14370
2024-04-27
Journal of Applied Clinical Medical Physics
Abstract:Purpose To evaluate the accuracy of different dosimeters and the treatment planning system (TPS) for assessing the skin dose due to the electron streaming effect (ESE) on a 1.5 T magnetic resonance (MR)‐linac. Method Skin dose due to the ESE on an MR‐linac (Unity, Elekta) was investigated using a solid water phantom rotated 45° in the x‐y plane (IEC61217) and centered at the isocenter. The phantom was irradiated with 1 × 1, 3 × 3, 5 × 5, 10 × 10, and 22 × 22 cm2 fields, gantry at 90°. Out‐of‐field doses (OFDs) deposited by electron streams generated at the entry and exit surface of the angled phantom were measured on the surface of solid water slabs placed ±20.0 cm from the isocenter along the x‐direction. A high‐resolution MOSkinTM detector served as a benchmark due to its shallower depth of measurement that matches the International Commission on Radiological Protection (ICRP) recommended depth for skin dose assessment (0.07 mm). MOSkinTM doses were compared to EBT3 film, OSLDs, a diamond detector, and the TPS where the experimental setup was modeled using two separate calculation parameters settings: a 0.1 cm dose grid with 0.2% statistical uncertainty (0.1 cm, 0.2%) and a 0.2 cm dose grid with 3.0% statistical uncertainty (0.2 cm, 3.0%). Results OSLD, film, the 0.1 cm, 0.2%, and 0.2 cm, 3.0% TPS ESE doses, underestimated skin doses measured by the MOSkinTM by as much as –75.3%, –7.0%, –24.7%, and –41.9%, respectively. Film results were most similar to MOSkinTM skin dose measurements. Conclusions These results show that electron streams can deposit significant doses outside the primary field and that dosimeter choice and TPS calculation settings greatly influence the reported readings. Due to the steep dose gradient of the ESE, EBT3 film remains the choice for accurate skin dose assessment in this challenging environment.
radiology, nuclear medicine & medical imaging