Continuous On-Line Monitoring of Extracellular Ascorbate Depletion in the Rat Striatum Induced by Global Ischemia with Carbon Nanotube-Modified Glassy Carbon Electrode Integrated into a Thin-Layer Radial Flow Cell

Mn Zhang,K Liu,Kp Gong,L Su,Y Chen,Lq Mao
DOI: https://doi.org/10.1021/ac051188d
IF: 7.4
2005-01-01
Analytical Chemistry
Abstract:This study describes a novel analytical system integrating in vivo microdialysis sampling with a radial thin-layer flow cell with a single-walled carbon nanotube (SWNT)-modified glassy carbon electrode as working electrode for continuous and on-line monitoring of ascorbate depletion in the rat striatum induced by global ischemia. The SWNTs, especially those after vacuum heat treatment at 500 degrees C, are found to be able to enhance the electron-transfer kinetics of ascorbate oxidation at a low potential (ca. -50 mV) and possess a strong ability against electrode fouling. These properties essentially make it possible to determine ascorbate with a good stability and high selectivity against catecholamines and their metabolites and other electroactive species of physiological levels. While being integrated with in vivo microdialysis to assemble an on-line analytical system, the electrode is proved useful for continuous and sensitive monitoring of the basal dialysate level of ascorbate and its depletion in the rat striatum induced by global ischemia. The basal dialysate level of ascorbate is determined to be 5.0 +/- 0.5 mu M (n = 5) and a 50 +/- 10% (n = 3) depletion is recorded for the basal ascorbate after 4 h of global ischemia.
What problem does this paper attempt to address?