A 21.6μW inductively powered implantable IC for blood flow measurement

pradeep basappa khannur,kok lim chan,jia hao cheong,kai kang,andreas astuti lee,xin liu,huey jen lim,k ramakrishna,minkyu je
DOI: https://doi.org/10.1109/ASSCC.2010.5716602
2010-01-01
Abstract:This paper presents a fully integrated inductively powered implantable circuits for blood flow measurement, which are embedded within vascular prosthetic grafts for early detection of graft degradation or failure. The ASIC interfaces with micro-fabricated pressure sensors and uses a 13.56MHz carrier frequency for power transfer and command/data communication. A backscatter-modulated passive telemetry is used for transmitting sensor readout information to an external monitoring device. The chip has been fabricated in 0.18μm CMOS process, occupies a total active area of 1.5×1.78mm2 and consumes a total power of 21.6μW. The rectifier achieves an efficiency of 66% The sub-μW 10-bit SAR ADC achieves an ENOB of 8.5 bits at 106KS/s conversion rate.
What problem does this paper attempt to address?