Tissue Vibration Pulsatility for Arterial Bleeding Detection Using Doppler Ultrasound

Zhiyong Xie,Eung-Hun Kim,Yongmin Kim
DOI: https://doi.org/10.1109/iembs.2009.5335088
2009-01-01
Abstract:Trauma is the number one cause of death among Americans between 1 and 44 years old, and exsanguination due to internal bleeding resulting from arterial injuries is a major factor in trauma deaths. We have evaluated the feasibility of using tissue vibration pulsatility in arterial bleeding detection. Eight femoral arteries from four juvenile pigs were punctured transcutaneously with a 6 or 9-French catheter. Also, 11 silicone vessels wrapped with turkey breast were placed in a pulsatile flow phantom and penetrated with an 18-gauge needle. The tissue vibration pulsatility was derived as a ratio of the maximum spectral energy from 200 to 2500 Hz of tissue vibration in systole over a baseline value in diastole. Then, the tissue vibration pulsatility index (TVPI) was defined as the maximum tissue vibration pulsatility value for each experimental condition. Both in vitro and in vivo results showed that the TVPI from injured vessels is significantly higher (p<0.005) than that of intact vessels. In addition, we constructed the 2D map of tissue vibration pulsatility during in vitro studies and found that it could be used for spatial localization of the puncture site. Our preliminary results indicate that the tissue vibration pulsatility may be useful for detecting arterial bleeding and localizing the bleeding site.