Calibration-free wavelength modulated TDLAS under high absorbance conditions.

Peng Zhimin,Ding Yanjun,Che Lu,Li Xiaohang,Zheng Kangjie
DOI: https://doi.org/10.1364/OE.19.023104
IF: 3.8
2011-01-01
Optics Express
Abstract:Currently, the method that uses a first-order Taylor series to approximate laser transmission has seriously affected the gas concentration measurement accuracy of tunable diode laser-absorption spectroscopy (TDLAS). This paper employs a second-order Taylor series to approximate laser transmission, and a high-precision second-order algorithm has been established that can determine the gas concentration directly. Then, this algorithm is used to test the NH3 mole fraction in a cell with NH3-Air mixtures. Experimental results show that the second-order algorithm not only effectively improves the measurement accuracy of gas concentration but also greatly broadens the scope of TDLAS. (C) 2011 Optical Society of America
What problem does this paper attempt to address?