Cooling Time Reduction for Focused Ultrasound Surgery by Dynamically Adjusting Sonication Time

Li Dehui,Shen Guofeng,Bai Jinfeng,Chen Yazhu
DOI: https://doi.org/10.1063/1.3607942
2011-01-01
Abstract:Cooling intervals are usually introduced between each sonication to avoid near-field heating in focused ultrasound surgery (FUS). However, this prolongs the treatment time greatly. A method of dynamically adjusting each sonication time with fewer cooling intervals is presented in this paper. With this method, cooling intervals were introduced between multiple consecutive sonications, not between individual sonications. To ensure the safety, consecutive sonication durations were unequal; usually the sequential durations were shorter. Two procedures, namely, a thermal dose (TD) feedback control procedure and a TD conformability optimization procedure, were utilized to determine the duration of each pulse. Firstly, several monitoring points were set on the single focus lesion boundary, mainly along the axial and radial directions. When the monitoring points reached the thermal necrosis threshold, the current pulse ended. Secondly, comparing the final total necrosis lesion with the planning target region (PTV), the conformal index was maximized by adjusting sonication durations. An 8x8x14 mm(3) target was simulated using the proposed method. The simulated results showed that the lesion formation conformed well with the PTV with 84% coverage index (CI) and 7.41% external volume index (EI). Moreover, the maximum temperature in the plane, 2 cm in front of focal plane, was 42.3 degrees C, thus demonstrating that the method could effectively prevent near-field heating. Ex vivo experiments were performed to validate the simulated results. Both simulated and experimental results proved that the proposed method could decrease the cooling time significantly, and was a safe and promising approach to FUS.
What problem does this paper attempt to address?