Inductively coupled plasma atomic emission spectrometric determination of acid-soluble metal elements chromium,manganese, zinc and copper in talcum powder

HU Xiao-jing,ZENG Ze,WANG Chang-wen,QIU Xin-yue,MU Ming-ren,FU Yao,SHEN Gui-ling,LIU Xiang-kuan
DOI: https://doi.org/10.3969/j.issn.1000-7571.2011.09.002
2011-01-01
Abstract:After digestion by aqua regia with microwave and selection of corresponding spectral lines at 267.7, 257.6, 213.8 and 324.7 nm as analytical lines, the acid-dissoluble metal elements including chromium, manganese, zinc and copper in talcum powder were determined by inductively coupled plasma atomic emission spectrometry (ICP-AES) under optimized conditions with radio frequency power of 1250 W, auxiliary air flow rate of 0.60 L/min and atomizer pressure of 26 psi. When the concentration of each testing element was 2.0 μg/mL, the interferences caused by 20.0 μg/mL of iron, magnesium, calcium, aluminum, zinc, nickel and copper were all less than 5%. Since the contents of these elements in talcum powder were less than 1%, their influence on the determination could be ignored. The detection limits of chromium, manganese, zinc and copper were 0.0048, 0.0038, 0.001 and 0.0026 μg/mL, respectively. This proposed method was applied in actual samples, and the determination results were consistent with those obtained by inductively coupled plasma mass spectrometry (ICP-MS) with relative standard deviations (RSD, n=8) of 1.1%-4.2% and recoveries of 93%-107%.
What problem does this paper attempt to address?