Broadband Superluminescent Diode-Based Ultrahigh Resolution Optical Coherence Tomography for Ophthalmic Imaging

Dexi Zhu,Meixiao Shen,Hong Jiang,Ming Li,Michael R. Wang,Yuhong Wang,Lili Ge,Jia Qu,Jianhua Wang
DOI: https://doi.org/10.1117/1.3660314
IF: 3.758
2011-01-01
Journal of Biomedical Optics
Abstract:Spectral domain optical coherence tomography (SD-OCT) with ultrahigh resolution can be used to measure precise structures in the context of ophthalmic imaging. We designed an ultrahigh resolution SD-OCT system based on broadband superluminescent diode (SLD) as the light source. An axial resolution of 2.2 μm in tissue, a scan depth of 1.48 mm, and a high sensitivity of 93 dB were achieved by the spectrometer designed. The ultrahigh-resolution SD-OCT system was employed to image the human cornea and retina with a cross-section image of 2048 × 2048 pixels. Our research demonstrated that ultrahigh -resolution SD-OCT can be achieved using broadband SLD in a simple way.
What problem does this paper attempt to address?