Development of Imaging System for Optical Coherence Tomography in Ophthalmology

HE Wei-hong,HE Yong-hong,LI Peng,ZHU Yi,SHAO Yong-hong,LIANG Rui-sheng,ZENG Nan,MA Hui
DOI: https://doi.org/10.3321/j.issn:1004-924x.2008.03.012
2008-01-01
Optics and Precision Engineering
Abstract:In order to obtain the Optical Coherence Tomography(OCT) images of macular region and optic nerve head in an eyeground,the application of OCT to ophthalmology is investigated.Because of the high transmittance of ocular media,an OCT system with Broadband Source(SLD) centered at 854.8 nm wavelength is applied for imaging in ophthalmology.Based on rapid scanning optical delay line in Fourier domain,the rapid scanning imaging is realized.Then,a device consisting of a OCT imaging path,a lighting path and a photographic path is developed to detect eyeground to obtain the OCT images of macular region and optic nerve head in the eyeground,and the system imaging depth is 3 mm.The experimental results show that OCT is a biomedical imaging method with high resolution,non-invasive and non-contact detecting for internal tissue structure and physiology.
What problem does this paper attempt to address?