Optimization and Calibration of Slat Position for a SPECT With Slit-Slat Collimator and Pixelated Detector Crystals

Xiao Deng,Tianyu Ma,Roger Lecomte,Rutao Yao
DOI: https://doi.org/10.1109/TNS.2011.2163418
IF: 1.703
2011-01-01
IEEE Transactions on Nuclear Science
Abstract:To expand the availability of SPECT for biomedical research, we developed a SPECT imaging system on an existing animal PET detector by adding a slit-slat collimator. As the detector crystals are pixelated, the relative slat-to-crystal position (SCP) in the axial direction affects the photon flux distribution onto the crystals. The accurate knowledge of SCP is important to the axial resolution and ...