Flow injection-spectrophotometry analysis of manganese in seawater samples

Sichao Feng,Zhigang Chen,Dongxing Yuan
2011-01-01
Abstract:A flow injection analysis (FIA) method coupled with spectrophotometry detection utilizing the complex reaction between 1-(2-pyridylazo)-2-naphthol and manganese was developed for the determination of dissolved manganese in seawater. A program written in labview was used to control the detecting system. A method detection limit (MDL) 14 nmol/L and linear range 0.020-2.000 μmol/L were obtained in ultra pure water matrix and possible interferences from Zn, Cu, Ni, Fe were also investigated. Because of the lack of low-Mn seawater as the matrix solution of the calibration curve and the FIA carrier, the MDL and linear range of the seawater detection still needs further confirmed. The results show that this method is easy to operate, fast and with wider linear range, and can be applied as on-field analytical technique for dissolved manganese in seawater.
What problem does this paper attempt to address?