Study of Detecting Depth Improvement for Open Electrical Impedance Tomography

何传红,何为,张莉,张占龙
DOI: https://doi.org/10.16182/j.cnki.joss.2011.09.021
2011-01-01
Abstract:Based on the analysis of critical factors and solutions of the detecting depth of open electrical impedance tomography(EIT),two-dimensional physical models and corresponding mathematical models for the human torso aiming at practical application were set up,and three excitation patterns,namely surface excitation(SE),back excitation(BE) and hand excitation(HE) were proposed.The validity of these presented excitation patterns was proved by simulation with finite element method(FEM).The simulation results indicate that the BE and HE make more excitation current distributed in deep region than SE,which lead to the boundary voltage is more sensitive to the internal impedance of object,then the detecting depth of open EIT is increased.Therefore,BE and HE are more suitable for application in open EIT.
What problem does this paper attempt to address?