Neutron Dose Measurement in Carbon Ion Radiation Therapy at HIRFL (IMP)

Su You-Wu,Li Zong-Qiang,Li Wu-Yuan,Wang Gui-Ling
DOI: https://doi.org/10.1088/1674-1137/34/5/006
2010-01-01
Abstract:For radiation protection purposes, the neutron dose in carbon ion radiation therapy at the HIRFL (Heavy Ion Research Facility in Lanzhou) was investigated. The neutron dose from primary C-12 ions with a specific energy of 100 MeV/u delivered from SSC was roughly measured with a standard Anderson-Broun rem-meter using a polyethylene target at various distances. The result shows that a maximum neutron dose contribution of 19 mSv in a typically surface tumor treatment was obtained, which is less than 1% of the planed heavy ion dose and is in reasonable agreement with other reports. Also the gamma-ray dose was measured in this experiment using a thermo luminescent detector.
What problem does this paper attempt to address?