Experimental Study of Interaction Between Radiofrequency Ablation and Realtime Temperature Monitoring

黄瑛,胡兵,申锷
DOI: https://doi.org/10.13929/j.1672-8475.2010.03.041
2010-01-01
Abstract:Objective To evaluate the interaction of radiofrequency ablation(RFA) and realtime temperature monitoring with a transparent tissue-mimicking phantom mainly composed of polyacrylamide gel.Methods The phantom,which mainly contained polyacrylamide gel and bovine serum albumin as a temperature sensitive indicator,was prepared with citrate buffer.Before the gel's solidification,urethral catheter was inserted into the gel solution until the solution solidified.The electrod was inserted into the phantom vertically,1 cm away from the urethral catheter.The coagulation lesions were evaluated with 2D sonography,T1 and T2 weighted imaging of MR,including the shapes,sizes,volumes and decenter ratios.Results The temperature curves showed tiny trembles when radifrequency was operated,but differences were not significant.The results showed the coagulation regions during temperature monitoring were smaller than those without temperature monitoring.Conclusion The catheter can realtime monitor the temperature precisely,and has mild impact of RFA.Under realtime monitoring of temperature,the coagulation region of RFA become smaller and more eccentric than those without temperature monitoring.
What problem does this paper attempt to address?