Comparison of diffraction enhanced imaging and in-line outline X-ray imaging with synchrotron radiation for mouse kidney

Yi-feng PENG,Yun-yan CHEN,Jing-li TIAN,Qi WANG,Guang-yu TANG,Ti-qiao XIAO,Pei-ping ZHU
DOI: https://doi.org/10.13929/j.1003-3289.2010.11.067
2010-01-01
Abstract:Objective: To evaluate the synchrotron radiation (SR) of X-ray phase contrast imaging on the microstructure of mouse kidneys, and to explore the imaging differences on kidney images obtained with diffraction enhanced imaging (DEI) and in-line outline X-ray imaging (IOXI) with SR facility. Methods: The mouse kidneys were taken as samples. DEI images were brought out at 4W1A beam line of Beijing Synchrotron Radiation Facility (BSRF), and IOXI images were recorded at BL13W beam line of Shanghai Synchrotron Radiation Facility (SSRF). Contrast of soft tissue and spatial resolution were selected mainly to evaluate the image quality. Results: The kidney microstructures were displayed clearly with both of the imaging modalities. However, more minute structures such as the renal glomeruli, were resolvable only in images taken with IOXI, the spatial resolution was about 10 μm. Conclusion: The phase contrast imaging has great potential to reveal the structures inside soft tissues, being able to achieve images of high quality for the tiny vessels. Able to obtain more information and having the sample setup and fine images, IOXI may be a better manner in study of biological tissue imaging.
What problem does this paper attempt to address?