FT-IR Based Loss-Spectrum Measuring System for Infrared Hollow Waveguides

Cong-Hui Yang,Hua,Wei Tan,Yi-Wei Shi,Katsumasa Iwai,Mitsunobu Miyagi
DOI: https://doi.org/10.1117/12.841916
2010-01-01
Abstract:A loss-spectrum measuring-system for hollow waveguides was established based on the Fourier Transform infrared Spectrometer (FTIR). In order to obtain better repeatability, we designed and fabricated two couplers for the system. One is a silver-coated hollow tube with the same inner diameter as the measured hollow waveguide. The other is a silver-coated tapered coupler whose inner diameter of the output end is the same as the measured hollow waveguide. Characteristics of the measuring system were discussed theoretically and experimentally when using the two couplers. Several kinds of hollow waveguides were measured and the properties of the loss spectra were discussed. The measured loss property showed that loss property depends on the output divergence angle of the coupler. Tapered coupler has a much larger output divergence than that of the hollow tube. Both of the couplers attains stable coupling with the hollow waveguides. Loss spectra were successfully measured for hollow waveguides with the length ranging from several centimeters to 3 meters.
What problem does this paper attempt to address?