Noise Control for Medical Oxygen Generator Based on Near Field Acoustical Holography Testing
Ding Wenjie,Jin Tao,Liu Jinling,Ma Yushan
DOI: https://doi.org/10.3969/j.issn.1004-6801.2011.02.024
2011-01-01
Abstract:In order to reduce operating noise from home oxygen generator, noise field data acquisition and sound field reconstruction of oxygen generator had been completed by using near field acoustic holography technology. According to visual information, such as sound pressure frequency spectrum, sound pressure radiation cloud maps, sound frequency spectrum, space velocity distribution maps, and time/frequency/sound pressure map, simple and effective noise control measures were selected for the sound source, acoustic transmission route with sound insulation enclosures, damping rubber coating, gypsum board sound absorption, and vibration pad for generator components. The practice shows that the near field acoustic holography technique is an effective investigation tool for the control of noise sources and sound radiation characteristics.