Design of a Low-Power Non-Invasive Glucose Monitor and Test of Its Performance

Yinzeng HE,Honghui XIAO,Lingqian CHANG,Chunxiu LIU,Xinxia CAI
DOI: https://doi.org/10.3969/j.issn.1004-1699.2010.07.002
2010-01-01
Abstract:Design of a low-power based non-invasive glucose monitor and the test of its performance are introduced. Based on reverse iontophoresis,the device can extract substances from subcutaneous space outward to the skin surface on-invasively. Low-oncentration glucose can be detected by a high-sensitively enzyme biosensor. In order to improve stability of the monitor and reduce the power consumption,microcontroller MSP430F1611 was used in this device. The system’s maximum working current is reduced to 17. 31mA at 3. 7 V voltage supply. The experiments results proved that the device can extract glucose from tissue liquid through skin and detect low concentration glucose. The system works in low frequency field,so skin can be simulated by resistance. The performance test proved that if resistance is less than 80 kΩ the extracting current is constancy 100 μA ,else the extracting voltage is constancy 8 V. Low concentration glucose can be obtained by reverse iontophoresis in pig experiment. In the experiment of low concentration glucose,good linearity of glucose concentration and responsible current is obtained between 5 ~ 200 μmol/L with the linear correlation coefficient of 0. 9979. The detection sensitivity of biosensor is 15. 619 nA( μmol/L) -1.