Phantom Study for Identifying Respiratory Motion and Correcting Motion Artifacts in SPECT Imaging

XU Quansheng,AN Jinggang,YUAN Kehong,YE Datian
DOI: https://doi.org/10.16511/j.cnki.qhdxxb.2010.11.027
2010-01-01
Abstract:Respiratory motion identification and corrections for the motion artifacts in SPECT imaging were analyzed in a phantom study by identifying,the blur direction in the SPECT image through a directional derivative analysis of the motion blurred image.The blur extent was estimated by an auto-correlation analysis of the pseudo-whitened blurred image.Then,a Wiener filter and the Richardson-Lucy blind deconvolution algorithm(RL) were employed to restore the motion blurred image according to the estimated point spread function.The results show that the identification results for 5 typical simulated motions are very close to the true value and that the RL algorithm enables better motion artifact correction within 5 iterations than the Wiener filter.This method can be used to identify respiratory motion with a post-reconstruction based deconvolution algorithm as a promising method in clinical practice to correct motion artifacts in SPECT/PET images.