Analysis of static force and fatigue between thread structure of dental implant and contact surface

Fei Gao,En-wei Zhang,Shi-cheng Wei,Jing Fang
DOI: https://doi.org/10.3969/j.issn.1673-8225.2010.30.006
2010-01-01
Abstract:BACKGROUND: Many factors affluence the stability of dental implant, the implant-bone interface, implant-abutment interface, as well as the thread structure of implant, play a key role in long term reliability. OBJECTIVE: To analyze static force and fatigue between thread structure of dental implant and contact surface, and to explore the potentially fatigue fracture of dental implant. METHODS: A three-dimensional model of dental implant and alveolar bone were established by using CAD software CATIA and Ansys Workbench, respectively. Then the static force and fatigue between dental implant and alveolar bone were analyzed after meshing the model, defining attribute, connections, and loading. RESULTS: A two-segment abutment implant system was constructed using CAD software CATIA. The implant system was characterized by platform transfer, hexagonal location, and taper seal, which represented structure of implants. The size of implant system was designed according to actual sizes and the assemblage was based on standard of mechanical design. Stress concentrated in screw and interface of dentalim plant and fatigue safety factor of the region was lower. These are potentially fatigue fracture of dental implant.
What problem does this paper attempt to address?