New Test and Analysis of Position-Sensitive-Silicon-Detector

Lang Feng,Yu-Cheng Ge,He Wang,Feng-Ying Fan,Rui Qiao,Fei Lu,Yu-Shou Song,Tao Zheng,Yan-Lin Ye
DOI: https://doi.org/10.1088/1674-1137/33/1/011
IF: 2.944
2009-01-01
Chinese Physics C
Abstract:We have tested and analyzed the properties of two-dimensional Position-Sensitive-silicon-Detector (PSD) with new integrated preamplifiers. The test demonstrates that the best position resolution for 5.5 MeV α particles is 1.7 mm (FWHM), and the best energy resolution is 2.1%, which are notably better than the previously reported results. A scaling formula is introduced to make the absolute position calibration.