Experimental Verification of Therapeutic Doses for the Superficially-Placed Tumor Radiotherapy with Heavy Ions at Hirfl

Liu Xin-Guo,Li Qiang,Wu Qing-Feng,Tao Jia-Jun,Jin Xiao-Dong
DOI: https://doi.org/10.1088/1674-1137/33/2/014
IF: 2.944
2009-01-01
Chinese Physics C
Abstract:Up to now, clinical trials of heavy-ion radiotherapy for superficially placed tumors have been carried out for six times and over 60 selected patients have been treated with 80-100 MeV/u carbon ions supplied by the Heavy Ton Research Facility in Lanzhou (HIRFL) at the Institute of Modern Physics, Chinese Academy of Sciences since November, 2006. A passive irradiation system and a dose optimization method for radiotherapy with carbon-ion beams have been developed. Experimental verification of longitudinally therapeutic dose distributions was conducted under the condition of simulating patient treatment in the therapy terminal at HIRFL, The measured depth-dose distributions basically coincide. with the expected ones. These results indicate that the. irradiation system and the dose optimization method are effective in the ongoing carbon-ion radiotherapy for shallow-seated turners at HIRFL.
What problem does this paper attempt to address?