Finite element analysis of custom endoprosthesis stem length in distal femur reconstruction

任翔,屠重棋
2009-01-01
Abstract:Based on CT images of the right femur of male healthy volunteer, five three-dimensional finite element models A, B, C, D, E were developed using different stem-length (150 mm, 160 mm, 170 mm, 180 mm, 190 mm) custom endoprostheses to reconstruct 40% defect of the distal femur. The corresponding osteotomy length of the femoral shaft was 160, 170, 180, 190 and 200 mm. Four times of body mass load corresponding to the normal walking gait cycle (3 km/h) was applied. The maximum stress on the bone recorded in the medial proximal femur dropped from 106.8 Mpa to 91.78 Mpa when the stem-length increased from 150 mm to 190 mm. While the maximum stress on the cement recorded in the top one fourth medial dropped from 27.2 Mpa to 19.06 Mpa. Compared to the bone and cement the stress on the stem was greater but had few changes. No bone fracture, cement damage or stem failure occurred in all models. Using custom endoprostheses to reconstruct defect of the distal femur has good biomechanical foundation. Short stems could induce concentrated stress in the bone and cement, easily leading to bone fracture and cement damage. Increasing the length of the intramedullary stem often increases the level of the stress-shielding on the bone.
What problem does this paper attempt to address?