Development of alternating magnetic field intensity measurement setup for cancer therapy

Changling Wang,Ruizhi Xu,Ning Gu
DOI: https://doi.org/10.3969/j.issn.1001-0505.2009.04.028
2009-01-01
Abstract:In order to measure the distribution of alternating magnetic field intensity, the alternating magnetic field intensity measurement setup based on DSP (digital signal processing) was designed. It is made up of three parts, field sensor, signal processing and collection system, and three-axis motion system. The sensor is a small loop where the magnetic field can be considered uniform in respect to the larger area of alternating magnetic field. The signals from sensor are converted into digital signals for the sequent processing. The spatial distribution of field intensity can be obtained by moving the sensor with the three-axis motion system. DSP F2812 is chosen as the central control unit due to its good capability of controlling motion as well as the relatively high sampling frequency of the A/D convertor. By the measurement of magnetic field distribution with this equipment, the optimal field area for hyperthermia can be found out, providing a good referring factor for clinicians in the magnetic fluid hyperthermia (MFH) cancer therapy.