Development of AC magnetic heating setup for cancer therapy with magnetic fluid hyperthermia

吴亚,孙剑飞,郭全忠,顾宁
DOI: https://doi.org/10.3321/j.issn:1001-0505.2004.06.017
2004-01-01
Abstract:The setup produces an AC (alternating current) magnetic field at the gap between the magnetic poles. Ferrite core in excitation winding enhances the magnetic field intensity, and bridge topology in circuit achieves high efficiency of power conversion. The frequency and magnetic field intensity are 25-120 kHz and 6-16 kA/m, respectively. Intensity measurement shows the uniformity of the magnetic field. The temperature increment of the magnetic nano-materials can be up to 28°C with magnetic heating under room temperature. The results suggest a way for clinical application in the magnetic fluid hyperthermia (MFH) cancer therapy.