In vitro biomechanical study of cervical compression staple and anterior constrained cervical plate using a single level fusion model of the cervical spine

沈莹莹,刘伟强,韩大庆,Guy R.Fogel
DOI: https://doi.org/10.3321/j.issn:1000-0054.2009.06.025
2009-01-01
Abstract:The biomechanical characteristics of a low stiffness cervical compressive staple (CCS) were compared to those of an anterior constrained cervical plate (ACCP) using a single level fusion model to understand CCS properties. 12 human specimens with a C4-5 fusion with an interbody graft were tested in an in vitro biomechanical testing system. The specimens were tested for 15° flexion and 20° extension with the load sharing characteristics studied for 50 N and 100 N axial compression loads. The results show that the CCS has similar properties to the ACCP for the single level fusions and that the CCS performs better with the proper load on the graft.
What problem does this paper attempt to address?