An immunoassay chip based on electrowetting on dielectrics

Liang Zhu,Xiongying Ye,FENG Yan-Ying,FENG Jin-Yang,WU Yan-Bin,ZHOU Zhao-Ying
DOI: https://doi.org/10.3321/j.issn:0253-3820.2009.03.033
IF: 1.193
2009-01-01
Chinese Journal of Analytical Chemistry
Abstract:An immunoassay chip based on electrowetting on dielectrics (EWOD) droplet control method was developed. Basing on the process of immunoassay, the immunoassay chip was designed, and fabricated using indium-tin-oxides(ITO) coated glass as substrate. We tested the controllability performance oil droplet actuating of the. chip. The experimental results of contact angles of the droplet were closer to the theoretical predictions within 100 V. The largest speed of the droplet movement was up to 3.75 mm/s, and the droplet manipulations were achieved by EWOD on the ITO-coated glasses. By the immunoreaction of rat IgG and goal anti-rat IgG oil the immunoassay chip, the possibility of the immunoreaction oil the immunoassay chip was validated. The volume of sample is 0.5 mu L and the detected time is 20 min. The defected range of the immunoreaction of rat IgG and goat anti-rat IgG is 0.1 - 20 mg/L.
What problem does this paper attempt to address?