A Study of Blood Sugar Meter Embedded in Mobile Phone

Mo Bu,Jun Lin
DOI: https://doi.org/10.1109/ICCA.2007.4376505
2008-01-01
Abstract:A new portable blood sugar meter embedded in mobile phone is proposed in this paper. This system adopts MSP430 single chip microcomputer (SCM) as central controlling and software platform based on Java mobile phone. It is an inexpensive, convenient and rapid way to obtain blood sugar level. Furthermore, users can transfer the measure results to doctors or secure server and get the feedback Rx by the mobile network in time. We develop the method of data processing and discuss the temperature compensation. According to our experience, the maximal measure error of this system is 7.84 %, which is satisfied with the American Diabetes Association (ADA) recommendation.
What problem does this paper attempt to address?