Determination of aesculin concentration in rabbit pasma by RP-HPLC with fluorescence detector

Xiu-ping CHEN,Yu-ying WANG,Miao WANG,Qin WU,Jing-han SHI,Guan-hua DU
DOI: https://doi.org/10.3321/j.issn:1001-1978.2007.04.031
2007-01-01
Abstract:Aim: To establish a HPLC-FLD method for determining the concentration of aesculin in rabbit plasma. Methods: The sample was separated with a C 18 column (150 mm x 4.6 mm, 5 μm) at 30°C with the mobile phase of methanol and water (74:26); A fluorescence detector was used (λex = 230 nm λem = 460 nm); The flow rate was 1 ml·min-1. Results: The retention time of aesculin was (3.620 ± 0.05) min; A good linear relationship of aesculin in the plasma samples at the concentration between 0.1 mg·L-1 and 100 mg·L-1 was observed (r = 0.997); The mean absolute recovery was 74.0%-82.6% and the lowest detective limit was 0.05 ng; Both the intra-day and inter-day precisions were less than 9.0%. After a single intravenous injection of aesculin (5 mg·kg-1) in rabbits, the T 1/2a and T1/2β were (0.196 ± 0.008)·h-1 and (6.70 ± 0.645)· h-1. Conclusions: The method was simple, sensitive and credible and could be used for the pharmacokinetic study of aesculin.