The Feasibility of Using Bioimpedance Data at Lower Frequency Range to Calculate Human Body Water Distribution

Xiao-Yan SUN,Tao WANG
DOI: https://doi.org/10.3969/j.issn.0258-8021.2007.02.006
2007-01-01
Abstract:Objective: To find out appropriate method to calculate human body composition by comparing two groups of fluid volume value calculated by body impedance analysis using data at lower frequency range and whole frequency range, respectively. Methods: The data of body impedance was collected by Hydra 4200 BIS. The two groups of fluid volume value were calculated: one was drawn from Cole-Cole model by fitting impedance data at lower frequency (5kHz-100kHz) using Matlab, and the other was drawn from the modified Cole-Cole model using the impedance data up to higher frequency range (5kHz-1000kHz), which was given by Hydra 4200. The consistency of the two methods was tested by regression and Bland-Altman plot. Results: The regression factor of ECW and ICW between the two methods was 0.9985 and 0.9683, respectively. The Bland-Altman plots showed that consistency of ECW and ICW between the two methods was acceptable. Conclusion: Using impedance data measured at lower frequency range to evaluate human hydration status is feasible.