Theoretical and Experimental Study of Enhanced Subharmonic Emission from Microbubbles with Chirp Excitation

Gong Yan-Jun,Zhang Dong,Xi Xiao-Yu,Gong Xiu-Fen,Liu Zheng
DOI: https://doi.org/10.7498/aps.56.7051
IF: 0.906
2007-01-01
Acta Physica Sinica
Abstract:Subharmonic contrast imaging provides improved ultrasound imaging quality by taking advantage of increased contrast to tissue ratio. This article presents a novel approach to enhance the subharmonic emission from microbubbles by using a chirp excitation technique. The chirp excitation pressure threshold and the sound pressure dependence of subharmonic are theoretically evaluated based on the Church's model. Measurements validate that the optimized bandwidth and time duration of the chirp excitation can improve subharmonic amplitude and reduce side-lobe level. Both theoretical and experimental results demonstrate that the chirp excitation can improve the amplitude up to 22 dB over the pulse excitation at the subharmonic frequency when comparing the coded excitation with traditional pulse excitation.