Rapid gas chromatogram determination of methane, organic acid in in vitro ruminal fermentation products

Wei-lian HU,Jia-kun WANG,Jian-min L(U),Yan-qiu GUO,Jian-xin LIU
DOI: https://doi.org/10.3321/j.issn:1008-9209.2006.02.020
2006-01-01
Abstract:A rapid determination method of ruminal fermentation products by gas chromatogram has been developed. By using the HP-INNOWAX, FID and nitrogen as the carrier, the optimal condition for methane measurement was: nitrogen pressure at 130 kPa, total and column flow rates at 30.2 and 1.7 mL·min-1, and the column, injection and detection temperatures at 80, 100 and 120°C, respectively. The optimal condition for acetate, propionate and butyrate measurement was: nitrogen pressure at 90 kPa, total and column flow rates at 37.2 and 0.67 mL·min-1, and column, injection and detection temperatures at 180, 200 and 220°C, respectively.