Energy Loss Correction for A Crystal Calorimeter
He Miao,Wang Yi-Fang,Bian Jian-Ming,Cao Guo-Fu,Deng Zi-Yan,He Kang-Lin,Huang Bin,Ji Xiao-Bin,Li Gang,Li Hai-Bo,Li Wei-Dong,Liu Chun-Xiu,Liu Huai-Min,Ma Qiu-Mei,Ma Xiang,Mao Ya-Jun,Mao Ze-Pu,Mo Xiao-Hu,Qiu Jin-Fa,Sun Sheng-Sen,Sun Yong-Zhao,Wang Ji-Ke,Wang Liang-Liang,Wen Shuo-Pin,Wu Ling-Hui,Xie Yu-Guang,Yang Ming,You Zheng-Yun,Yu Guo-Wei,Yuan Chang-Zheng,Yuan Ye,Zang Shi-Lei,Zhang Chang-Chun,Zhang Jian-Yong,Zhang Ling,Zhang Xue-Yao,Zhang Yao,Zheng Zhi-Peng,Zhu Yong-Sheng,Zou Jia-Heng
DOI: https://doi.org/10.1088/1674-1137/32/4/006
IF: 2.944
2008-01-01
Chinese Physics C
Abstract:Material effect of inner-detectors on the performances of the BESIII Electromagnetic Calorimeter (EMC) is investigated. The BESIII Time-Of-Flight counters (TOF) have been utilized to improve the energy resolution and detection efficiency for photons after a careful energy calibration. A matching algorithm between TOF and EMC energy deposits is developed, and the effects of beam-related background are discussed. The energy resolution is improved and the photon detection efficiency can be increased by the combined measurement of EMC and TOF detectors.