Application of Second Derivative Spectrum Prepares in Quantification Measuring Glucose-6-phosphate and Fructose-6-phosphate Using a FTIR/ATR Method

CHEN Jie-mei,PAN Tao,CHEN Xing-dan
DOI: https://doi.org/10.3321/j.issn:1004-924x.2006.01.001
2006-01-01
Optics and Precision Engineering
Abstract:A simultaneous quantification measurement method of glucose-6-phosphate(G6P) and fructose-6-phosphate(F6P) was developed using a Fourier transform infrared(FTIR) spectrometer and attenuated total reflection(ATR) techniques.For applying the quantification method to the enzyme reaction system from G6P to F6P with phosphoglucose isomerase(PGI),the G6P-F6P-Tris mixture solution was used as the quantification object,and the spectra of the G6P-F6P-Tris mixture solutions which the concentrations were known were used as calibration spectra.Corresponding to the absorbance spectra or the second derivative spectra,selecting the value in main absorption peaks of 1 086 cm~(-1),(1 082 cm~(-1)),1 065 cm~(-1),1 036 cm~(-1) and 978 cm~(-1) or the data in the fingerprint region of 1 200~900 cm~(-1),four types of the multiple linear regressions analyses were performed.By comparing the related coefficients,the average relative error and the root mean square error cross validation(RMSECV) values for the calculated values and the actual values for the concentrations corresponding to all types,the enough accuracy could be achieved for the calculated values based on the data in the fingerprint region for the second derivative spectra.The calculated values corresponding to the second derivative spectra have the higher accuracy than ones corresponding to the absorbance spectra,and the calculated values based on the data in the fingerprint region have the higher accuracy than ones based on the data in main absorption peaks.
What problem does this paper attempt to address?