Combination Models of Absorption Peaks for FTIR/ATR Spectroscopy Analysis of Glucose Aqueous Solutions

WU Ling-ling,PAN Tao,XIE Jun
DOI: https://doi.org/10.3969/j.issn.1008-1402.2011.03.012
2011-01-01
Abstract:By using Fourier transform infrared spectroscopy(FTIR) and attenuated total reflection(ATR) technology,the analysis models for glucose aqueous solution was established.Based on the subtracted spectra of glucose aqueous solution for de-ionized water,five absorption peaks 1150,1103,1078,1034,991(cm-1) were found.Multiple linear regression(MLR) models of absorption peaks combination were established,the optimal absorption peaks combination was 1103,1034,991(cm-1),the corresponding root mean squared error of predication(RMSEP),correlation coefficient of predication(RP) were 0.993 mmol/L,0.989 respectively.It was greatly better than one of partial least squares(PLS) model in the whole spectral collecting region(4500-600 cm-1),and model complexity was reduced largely.To get the stable and reliable results,all models were obtained by the average data based on 50 different divisions of calibration set and prediction set.
What problem does this paper attempt to address?