A Novel Method for Evaluating and Improving the ~1H-MRSI Glioma Data Quality

KH Yuan,HY Lu,SL Bao,QS Chen,SW Li,CJ Duan
DOI: https://doi.org/10.1080/10020070512331342890
2005-01-01
Progress in Natural Science
Abstract:Metabolic information obtained by proton magnetic resonance spectroscopic imaging (H-1-MRSI) has been approved to be a powerful tool to identify either benign or malignant glioma, as well as to confirm the tumor level. However, H-1-MRSI data are affected by various factors, such as the thermal noise, eddy currents, susceptibility artifacts, and rigid body motion. To get accurate quantitative metabolic information, the key problem is to assess the H-1-MRSI data quality. In this paper, we introduce a new evaluating system to filter the data, and a new method, called wavelet denoising method, to improve the data quality under the evaluating system. Experimental results on H-1-MRSI glioma data demonstrate that preprocessing is prerequisite and the proposed algorithm with evaluating system is effective.
What problem does this paper attempt to address?