Experimental Study on Refractive Index and Concentration Measurement of Liquid by Confocal Concave Spherical Mirror Cavity F-P Interferometer
XIANG Wang-hua,ZHANG Bing,LIANG jie,ZHOU Yu,JIA Guang-ming,ZHANG Gui-zhong,LIU Guo-biao,LIU Chun-jiang
DOI: https://doi.org/10.3969/j.issn.0493-2137.2007.01.016
2007-01-01
Abstract:Based on confocal spherical Fabry-Perot (F-P) interferometer, a sensing device was designed for measuring refractive index and concentration of aqueous solution, which has a single-mode silica optical fiber for laser transmission and a two-unit CCD for image acquisition. It was manifested by a successful measurement of the refractive index of ethanol solution, with high precision. The measurement principle and feasibility of the device were analyzed, the device was constructed with a He-Ne laser, a segment of fiber, an F-P cavity and a CCD head. Through counting the number of the concentric interference rings and performing some mathematics, the re-fractive index and the solution concentration were retrieved with high precision. The dynamic evolution of these rings were demonstrated for the first time, corroborating the measurement process. A refractive index change of 10 -5 and a corresponding concentration variation of 1.0 ?10-5 could be inferred from the measured image data. Compared with traditional methods, this method is easy, simple, and capable of on-line monitoring.