High-Resolution Frequency Standard Comparator Based On A Special Phase Comparison Approach

H Zhou,W Zhou,Zq Xuan,H Wang,Cs Liu
DOI: https://doi.org/10.1109/freq.2004.1418546
2005-01-01
Abstract:Along with the accuracy enhancement of atomic frequency standards, the higher frequency standard comparison resolution is necessary. Based on the concept of the greatest common factor frequency and a common frequency source, utilizing phase coincidence detection approach the high-resolution phase comparison between frequency standards can be completed. The resolution of a simple instrument is better than 10 ps, and the resolution of theoretical analysis is much higher than it. In the long-term V frequency standard comparison higher than 1x10(-14) / hour accuracy can be obtained and for one day comparison the measuring precision can be 10(-16). This method can also be used in the measurement of short-term frequency stability or as a terminal display, device of frequency standard comparator, and 10(-13) /tau stability resolution can be obtained. In the practical device the phase coincidences between a high-stability common crystal oscillator signal f(c) and two compared frequency standard signals f(1) and are detected respectively, with the phase coincidence between and as a start signal and with the phase coincidence between f(c) and f(2) as a stop signal to build a gate time. This gate is synchronized by the multiple periods of the f(c), and +/- 1 count error can be eliminated when using f(c) as the counting signal. With this way, sampling the phase difference and measuring it highly accurately are combined together, higher resolution and low cost are easily obtained.
What problem does this paper attempt to address?