Accuracy of three-dimensional color Doppler ultrasonography in evaluating the extent of carotid arterial stenosis

邢晋放,曹铁生,段云友
DOI: https://doi.org/10.3321/j.issn:1673-8225.2004.12.028
2004-01-01
Abstract:Aim: To explore the accuracy of three-dimensional color Doppler ultrasonography in evaluating the extent of carotid artery stenosis. Methods: Forty patients with carotid artery stenosis were examined by using Acuson X/P 128 color Doppler ultrasonography diagnosis system and TomTec three-dimension color Doppler ultrasonography system. The indexes measured were volume of pathological changes (Vp) extent of stenosis volume(V%), maximal extent of stenosis area(S%) and maximal extent of stenosis diameter(D%). The results of different times for the same examiner, and of different physicians were analyzed by correlation, P < 0.05 meant statistical signification. Results: The range of V% was 26% - 80%, S% was 33% - 86%, and D% was 36% - 93%. High correlation was found in the results of different times for the same examiner (r = 0.94 - 0.97, P < 0.05), and that of different examiners (r = 0.88 - 0.93, P < 0.05). Conclusion: The extent of carotid artery stenosis can be quantitatively assessed exactly by three-dimensional color Doppler ultrasonography.
What problem does this paper attempt to address?