Spectra-Resolved Technique of A Sensitive Time-Resolved Fluorescence Immunoassay Instrument

ZY Guo,Z Tian,YL Jia
DOI: https://doi.org/10.1117/12.572442
2004-01-01
Abstract:The lanthanide trivalence ion and its chelates are used for marking substance in time-resolved fluorescence immunoassay (TRFIA), marking the protein, hormone, antibody, nucleic acid probe or biologic alive cell, to measure the concentration of the analysis substance inside the reaction system with time-resolved fluorometry after the reaction system occurred, and attain the quantitative analysis's purpose. TRFIA has been become a kind of new and more sensitive measure method after radioisotope marking, enzymatic marking, chemiluminescence, electrochemiluminescence, it primarily is decided by the special physics and chemistry characteristic of lanthanide trivalence ion and its chelates. In this paper, the result of spectroscopic evaluation of europium trivalence ion and its chelate, and the principle of spectra-resolved technology and a sensitive time-resolved fluorescence immunoassay instrument made by ourselves are reported. In the set, a high frequency Xenon pulsed-light was adopted as exciting light, and two special filters was utilized according to spectra-resolved technique. Thus the influence of scattering light and short-lifetime fluorescence was removed. And the sensitivity is 10(-12)mol/L (when Eu3+ was used for marking substance), examination repeat is CVless than or equal to5 %, examination linearity is from 10 (-8) mol/L to 10(-12)mol/L, correlation coefficient rgreater than or equal to95%(p<0.01).
What problem does this paper attempt to address?