Time-Resolved Fluorescence Immunoassay with Measurement of A Europium Chelate in Solution - Dissociation Conditions and Application for Determination of Cortisol

XD YANG,YX CI,WB CHANG
DOI: https://doi.org/10.1021/ac00087a026
IF: 7.4
1994-01-01
Analytical Chemistry
Abstract:A time-resolved fluorescence immunoassay method based on the quantification of the new europium chelator, ((chlorosulfonyl)thenoyl)trifluoroacetone (CTTA), in the presence of excess EU(3+) in solution with a novel detection principle is described. CTTA-labeled solid-phase immunocomplexes obtained by conventional immunoassay procedures were transferred into solution with a dissociation solution, and then fluorescent complexes were formed. The dissociating solution was optimized and consisted of sodium dodecyl sulfate, Eu3+, and tri-n-octylphosphine oxide, which acted as dissociating reagent, fluorescent lanthanide ion, and enhancement reagent, respectively. The new dissociation solution was very simple and available, and it was applied to a competitive immunoassay of cortisol, which had a detection limit of 0.5 ng/mL of cortisol with a dynamic range up to 2500 ng/mL. The average recovery in urine was 105%.
What problem does this paper attempt to address?