Chiral Analysis of Amino Acids in Biological Samples by Micellar Electrokinetic Chromatography with Laser-induced Fluorescence Detection

YR Fu,LS Yan,G Luo,CZ Chen,YM Wang,LT Zhang
DOI: https://doi.org/10.3321/j.issn:0253-3820.2004.12.003
2004-01-01
Abstract:A micellar electrokinetic chromatography (MEKC) with laser-induced fluorescence detection (LIF) for chiral analysis of amino acids in biological samples was established. These amino acids, including D or L-alanine (Ala), D or L-glutamic acid (Glu) and D or L-aspartic acid (Asp), were derivatized with fluorescein isothiocyanate (FITC). The optimized conditions in electrophoresis were that uncoated fused-silica capillary (50 mum x 37 cm, 30 cm to detector) was used, running voltage was 20 kV, running buffer was 40 mmol/L borate buffer (pH 9.3) containing 50 mmol/L sodium dodecyl sulfate (SDS) and 13 mmol/L beta-cyclodextrin (beta-CD), the sample solution was injected 5 s with pressure mode, separated at 25degreesC, and monitored with a LIF detection. The baseline separation of these amino acid enantiomers was achieved. The linear ranges of six FITC-derivatives was about 0.050 similar to 2.000 mg/L, and the determination limits (S/N = 3) were 0.005 similar to 0.050 mg/L. The average recoveries (n = 6) and RSD of the six FITC-derivatives were 96.33% similar to 102.0% and 3.6% similar to 5.2% from rat serum, 97.59% similar to 102.8% and 3.5% similar to 5.9% from rat lens. The within-day precisions (n = 9) and between-day precisions (n = 5) of them were lower than 5.0% and 6.5% from rat serum, lower than 5.0% and 5.5% from rat lens, respectively.
What problem does this paper attempt to address?