Electrochemical depletion of ascorbic acid in the detection of hydrogen peroxide - An investigation using SECM

Kang Wang,Hui Wei,Xinghua Xia
DOI: https://doi.org/10.3321/j.issn:0567-7351.2004.14.014
2004-01-01
Acta Chimica Sinica
Abstract:An electrochemical method for detection of hydrogen peroxide has been advocated via selective elimination of interferences (i. e., ascorbic acid). In this method, the elimination of the interferences was achieved by positioning a tip electrode of scanning electrochemical microscopy (SECM) in the diffusion layer of an Au electrode and meanwhile applying a suitable potential on the substrate. Influence of the tip-substrate distance and the potentials of tip- and substrate-electrode on the selectivity and sensitivity of hydrogen peroxide detection was systematically investigated. Results showed that highly selective and sensitive detection of hydrogen peroxide free from interference of ascorbic acid (0.05 mmol·L -1) could be obtained when the tip-substrate distance was 22.5 μm and the tip and substrate potentials were 0.5 and 0.4 V, respectively. The linear range for hydrogen peroxide detection is 4 × 10 -5 - 1 × 10 -3 mol · L -1. Finally, potential application of the present method in fabrication of micro-electrochemical sensing devices has been discussed.
What problem does this paper attempt to address?