Amperometric Biosensor for Hydrogen Peroxide Based on Electrodeposited Sub-micrometer Gold Modified Glassy Carbon Electrode

SQ Wang,J Chen,XQ Lin
DOI: https://doi.org/10.1002/cjoc.20040220409
IF: 5.56
2004-01-01
Chinese Journal of Chemistry
Abstract:A new type of hydrogen peroxide amperometric biosensor was fabricated based on electrochemically deposited sub-micrometer Au particles (sm-Au) on a glassy carbon electrode (GCE). Electrochemical deposition condition was optimized for obtaining uniformly distributed sub-micrometer sized Au array on the electrode surface. The hydrogen peroxide sensor was fabricated by adsorbing phenothiazine methylene blue (MB) molecules on the surface of sm-Au and covering a cross-linked horseradish peroxidase (HRP) layer, labeled as HRP/MB/sm-Au/GCE. The characteristics of this biosensor were evaluated with respect to applied potential and pH. The amperometric response of the sensor was linear to the H2O2 concentration over a wide range of 9.9×10^-6-1.11×10^-2 mol/L. A detection limit (s/n=3) of 3.0×10^-6 mol/L H2O2 was estimated for a sampled chronoamperometric detection at 1.5 min after potential step of 200 to -400mV vs. SCE. The immobilized MB molecules shuttled electrons at α= 0.77 and an apparent electron transfer rate constant of ks^0=0.053s^-1. Interference of ascorbic acid, dopamine and uric acid was investigated. This sensor has very good stability and reproducibility for long-term use.