Pulsed magnet system with liquid nitrogen cooling for direct measurement of magnetocaloric effect

Yongbo Tang,Yungui Chen,Hao Fu.,Baohua Teng,Huanxing Li,Baomu Wang,Qinxiu Xue,Mingjing Tu
DOI: https://doi.org/10.3321/j.issn:1002-185x.2003.01.018
2003-01-01
Rare Metal Materials and Engineering
Abstract:The pulsed magnet, described in. this paper operates at liquid nitrogen temperature and is used for the direct measurement of magnetocaloric effect of magnetic materials. It uses a relatively small pulsed power supply (22 kW). With pulsed duration of 1 2,3,4 and 5 seconds, respectively. The solenoid. has a bore diameter of 20 mm. The magnet system has operated steadily to 2. 232 T at 180 A in experiments. The uniformity of magnetic field-is better than 1% within 10 mm of solenoid center. It is satisfied for the testing of magnetocaloric effect of magnetic materials.
What problem does this paper attempt to address?