Design of ECT high speed data acquisition system for two-phase flow measurement

王雷,王保良,冀海峰,黄志尧,李海青
DOI: https://doi.org/10.3785/j.issn.1008-973X.2002.05.001
2002-01-01
Abstract:The filter in Electrical Capacitance Topography (ECT) system is a bottleneck which hampers the increase of sampling speed. Based on the charge amplification principle, a new ECT data acquisition circuit without filter was developed for two-phase flow measurement. This circuit has good stray-immunity for measuring small capacitance. It charges and discharges the detection capacitance only once during one measurement cycle. Compared with charge/discharge and AC-based circuit, this circuit doesn't need filter because there is no pulsating noise. Optimization step of parallel measurement, high speed CMOS chip and fast communication were adopted. Test results showed that the sensitivity of 4.8 V/pF, the resolution of 0.5 fF and the data capture rate of 600 frames/s can be achieved. It satisfies the requirements for industrial online operation.
What problem does this paper attempt to address?